Article,

Endoleak as a complication of endoluminal grafting of abdominal aortic aneurysms: classification, incidence, diagnosis, and management.

, , , , and .
J Endovasc Surg, 4 (2): 152--168 (May 1997)

Abstract

The inability to obtain or maintain a secure seal between a vessel wall and a transluminally implanted intra-aneurysmal graft is a complication unique to the evolving technique of endovascular aneurysm exclusion. Because the term "leak" has long been associated with aneurysm rupture, the term "endoleak" is proposed as a more definitive description of this phenomenon. Embracing both persistent blood flow into the aneurysmal sac from within or around the graft (graft related) and from patent collateral arteries (nongraft related), endoleak can be classified as primary or secondary depending on the time of occurrence (within 30 days of implantation or following apparent initial seal, respectively). Diagnostic techniques to detect endoleak include arteriography, intraprocedural pressure monitoring, contrast-enhanced computed tomography, abdominal X ray, and duplex scanning. Management strategies for endoleak range from observation with periodic imaging surveillance to correction by additional endoluminal or surgical procedures. Standardization of the terminology describing this important sequela to endovascular aneurysm exclusion should facilitate uniform reporting of clinical trial data vital to the evaluation of this emerging technique.

Tags

Users

  • @jmaiora