Objective
To assess the inter-rater reliability of MScanFit MUNE using a “Round Robin” research design.
Methods
Twelve raters from different centres examined six healthy study participants over two days. Median, ulnar and common peroneal nerves were stimulated, and compound muscle action potential (CMAP)-scans were recorded from abductor pollicis brevis (APB), abductor digiti minimi (ADM) and anterior tibial (TA) muscles respectively. From this we calculated the Motor Unit Number Estimation (MUNE) and “A50”, a motor unit size parameter. As statistical analysis we used the measures Limits of Agreement (LOA) and Coefficient of Variation (COV). Study participants scored their perception of pain from the examinations on a rating scale from 0 (no pain) to 10 (unbearable pain).
Results
Before this study, 41.6\% of the raters had performed MScanFit less than five times. The mean MUNE-values were: 99.6 (APB), 131.4 (ADM) and 126.2 (TA), with LOA: 19.5 (APB), 29.8 (ADM) and 20.7 (TA), and COV: 13.4 (APB), 6.3 (ADM) and 5.6 (TA). MUNE-values correlated to CMAP max amplitudes (R2-values were: 0.463 (APB) (p\textless0.001), 0.421 (ADM) (p\textless0.001) and 0.645 (TA) (p\textless0.001)). The average perception of pain was 4.
Discussion
MScanFit indicates a high level of inter-rater reliability, even with only limited rater experience and is overall reasonably well tolerated by patients. These results may indicate MScanFit as a reliable MUNE method with potential as a biomarker in drug trials.
%0 Journal Article
%1 sorensen_assessing_2021
%A Sørensen, Daniel Mosgaard
%A Bostock, Hugh
%A Ballegaard, Martin
%A Fuglsang-Frederiksen, Anders
%A Graffe, Carolina Canillo
%A Grötting, Arnstein
%A Jones, Kelvin
%A Kallio, Mika
%A Krarup, Christian
%A Krøigård, Thomas
%A Lupescu, Tudor
%A Maitland, Stuart
%A Moldovan, Mihai
%A Nilsen, Kristian Bernhard
%A Pugdahl, Kirsten
%A Santos, Miguel Oliveira
%A Themistocleous, Andreas Constantinos
%A Zlateva, Suzana Slavica
%A Ööpik, Merle
%A Tankisi, Hatice
%D 2021
%J Neurophysiologie Clinique
%K Disease Inter-rater MScanFit MUNE, Motor Neurodegenerative disease, estimation, neurone number unit variability,
%R 10.1016/j.neucli.2021.11.002
%T Assessing inter-rater reproducibility in MScanFit MUNE in a 6-subject, 12-rater “Round Robin” setup
%U https://www.sciencedirect.com/science/article/pii/S0987705321001234
%X Objective
To assess the inter-rater reliability of MScanFit MUNE using a “Round Robin” research design.
Methods
Twelve raters from different centres examined six healthy study participants over two days. Median, ulnar and common peroneal nerves were stimulated, and compound muscle action potential (CMAP)-scans were recorded from abductor pollicis brevis (APB), abductor digiti minimi (ADM) and anterior tibial (TA) muscles respectively. From this we calculated the Motor Unit Number Estimation (MUNE) and “A50”, a motor unit size parameter. As statistical analysis we used the measures Limits of Agreement (LOA) and Coefficient of Variation (COV). Study participants scored their perception of pain from the examinations on a rating scale from 0 (no pain) to 10 (unbearable pain).
Results
Before this study, 41.6\% of the raters had performed MScanFit less than five times. The mean MUNE-values were: 99.6 (APB), 131.4 (ADM) and 126.2 (TA), with LOA: 19.5 (APB), 29.8 (ADM) and 20.7 (TA), and COV: 13.4 (APB), 6.3 (ADM) and 5.6 (TA). MUNE-values correlated to CMAP max amplitudes (R2-values were: 0.463 (APB) (p\textless0.001), 0.421 (ADM) (p\textless0.001) and 0.645 (TA) (p\textless0.001)). The average perception of pain was 4.
Discussion
MScanFit indicates a high level of inter-rater reliability, even with only limited rater experience and is overall reasonably well tolerated by patients. These results may indicate MScanFit as a reliable MUNE method with potential as a biomarker in drug trials.
@article{sorensen_assessing_2021,
abstract = {Objective
To assess the inter-rater reliability of MScanFit MUNE using a “Round Robin” research design.
Methods
Twelve raters from different centres examined six healthy study participants over two days. Median, ulnar and common peroneal nerves were stimulated, and compound muscle action potential (CMAP)-scans were recorded from abductor pollicis brevis (APB), abductor digiti minimi (ADM) and anterior tibial (TA) muscles respectively. From this we calculated the Motor Unit Number Estimation (MUNE) and “A50”, a motor unit size parameter. As statistical analysis we used the measures Limits of Agreement (LOA) and Coefficient of Variation (COV). Study participants scored their perception of pain from the examinations on a rating scale from 0 (no pain) to 10 (unbearable pain).
Results
Before this study, 41.6\% of the raters had performed MScanFit less than five times. The mean MUNE-values were: 99.6 (APB), 131.4 (ADM) and 126.2 (TA), with LOA: 19.5 (APB), 29.8 (ADM) and 20.7 (TA), and COV: 13.4 (APB), 6.3 (ADM) and 5.6 (TA). MUNE-values correlated to CMAP max amplitudes (R2-values were: 0.463 (APB) (p{\textless}0.001), 0.421 (ADM) (p{\textless}0.001) and 0.645 (TA) (p{\textless}0.001)). The average perception of pain was 4.
Discussion
MScanFit indicates a high level of inter-rater reliability, even with only limited rater experience and is overall reasonably well tolerated by patients. These results may indicate MScanFit as a reliable MUNE method with potential as a biomarker in drug trials.},
added-at = {2023-02-08T12:07:40.000+0100},
author = {Sørensen, Daniel Mosgaard and Bostock, Hugh and Ballegaard, Martin and Fuglsang-Frederiksen, Anders and Graffe, Carolina Canillo and Grötting, Arnstein and Jones, Kelvin and Kallio, Mika and Krarup, Christian and Krøigård, Thomas and Lupescu, Tudor and Maitland, Stuart and Moldovan, Mihai and Nilsen, Kristian Bernhard and Pugdahl, Kirsten and Santos, Miguel Oliveira and Themistocleous, Andreas Constantinos and Zlateva, Suzana Slavica and Ööpik, Merle and Tankisi, Hatice},
biburl = {https://www.bibsonomy.org/bibtex/2c29e058ab3a2ac38361cd6b4a675b829/stumaitland},
doi = {10.1016/j.neucli.2021.11.002},
interhash = {e04f97b669a20d8054fd3db5b01c477a},
intrahash = {c29e058ab3a2ac38361cd6b4a675b829},
issn = {0987-7053},
journal = {Neurophysiologie Clinique},
keywords = {Disease Inter-rater MScanFit MUNE, Motor Neurodegenerative disease, estimation, neurone number unit variability,},
month = dec,
timestamp = {2023-02-08T12:08:14.000+0100},
title = {Assessing inter-rater reproducibility in {MScanFit} {MUNE} in a 6-subject, 12-rater “{Round} {Robin}” setup},
url = {https://www.sciencedirect.com/science/article/pii/S0987705321001234},
year = 2021
}