Abstract
This paper reports on validation experiments with the recently developed microAeth�, a pocket-sized device which is able to obtain real-time and personal measurements of black carbon (BC) aerosol. High reproducibility was observed when comparing the results from six new individual units during fixed-site monitoring out of a window (relative standard deviation RSD = 8% � 5%, N = 1442). The results obtained from the microAeth devices agreed with those obtained from a full size rack mounted Aethalometer, based on both the 1-minute data (R = 0.92, slope = 1.01 � 0.01, N = 1380) and 24-h average data. The 24-h average of real time data obtained from the microAeths was comparable to the BC concentration obtained from 24-h integrated PM2.5 filter deposits, as determined by multi-wavelength optical absorption (R = 0.98, slope = 0.92 � 0.07, N = 12). Rapid environmental changes in relative humidity (RH) and temperature (T) can result in false positive and negative peaks in the real time BC concentrations, though averages > 1-2-hour are only minimally affected. An inlet with a diffusion drier based on Nafion� tubing was developed in order to use BC data with a high time resolution. The data shows that the diffusion drier greatly reduce the impacts from rapid changes in RH and T when the monitoring system is worn in close proximity to the body (e.g., in the vest pocket).
Users
Please
log in to take part in the discussion (add own reviews or comments).